MODEL 4200/KEW 4202DIGITAL EARTH CLAMP TESTERINSTRUCTION MANUALKYORITSU ELECTRICAL INSTRUMENTSWORKS,LTD.MODEL4200 KEW4202
̶ 8 ̶M-4102ABy applying the Voltage (V) to the circuit from the Transformer jaw (CT1), current I is (shall be flowed) flowed corresponding to the eart
̶ 9 ̶ 6. Getting started(1) Check the battery voltage When nothing is displayed on the LCD, press the power button and turn on the instrument. B
̶ 10 ̶ 7. Measuring method#DANGER Never make measurement on a circuit in which the electrical potential exceeds AC300V. Transformer jaw
̶ 11 ̶7-1 Normal measurement of current * Press the Function button and select the ACA function. * Confirm the displayed unit is " mA "
̶ 12 ̶#CAUTION Follow the procedure described in "7-1 Normal measurement of current" and measure the current flowing on the earthed wi
̶ 13 ̶ 8. Other functions8-1 Auto power-off function This function is to prevent the instrument from being left turned on and conserve battery power
̶ 14 ̶8-5 Memory function This function is to save and display the measurement results. Saving the measurement results (1) Select any data numb
̶ 15 ̶8-6 Bluetooth Communication Function (KEW4202 only) KEW4202 can perform data communication between Android devices via Bluetooth commu
̶ 16 ̶Set-up: First, set a Bluetooth connection (pairing) from an Android device. For further details, please refer to the instruction man
̶ 17 ̶ 9. Battery replacement#WARNING In order to avoid possible shock hazard, take off the instrument from the conductor under test and
Contents 1. Safety warnings ... 1 2. Features ...
̶ 18 ̶ 10. Pairing with Bluetooth compatible devices (KEW4202)It is necessary to perform the pairing procedure to connect KEW4202 and Android devices
̶ 19 ̶ 11. Features of KEW Smart (KEW4202)Remote checking of measurements is possible without accessing KEW4202 using the special Android application
̶ 20 ̶MEMO
̶ 21 ̶MEMO
4-12 92-2088DISTRIBUTORKyoritsu reserves the rights to change specifications or designs described in this manual without notice and without ob
̶ 1 ̶ 1. Safety WarningsThis instrument has been designed, manufactured and tested according to IEC 61010: Safety requirements for Elect
̶ 2 ̶#DANGER Never make measurement on a circuit in which the electrical potential exceeds AC300V. Do not make measurement when thunder is ru
̶ 3 ̶ 2. FeaturesThis instrument is a digital clamp-on earth resistance tester, and it is used in multi-earthed systems. Can measure the earth resista
̶ 4 ̶ 3. Specification Operating system Earth resistance function: Constant voltage injection, Current detection, (Fr
̶ 5 ̶ Auto power-off Turns power off about 10 minutes after the last button operation. Applicable standards IEC61010-1 (CAT. IV 300V Pollu
̶ 6 ̶ 4. Instrument layout Symbols to be displayed on the LCD1263457891110 Name of each parts and buttons (for MODEL4200 and KEW4202) 1 Transforme
̶ 7 ̶ 5. Measurement principleThis instrument can measure the earth resistance to earth in multi-earthed system.Let's regard earth resistance und