3.3 KEW6305
3.1.2 AC Power supply
Check the followings before connecting the Power cord.
DANGER
Use only the Power cord supplied with this instrument.
Connect the Power cord mains plug to a mains socket outlet. The mains supply voltage must not
exceed AC240V. (max rated voltage of supplied Power cord MODEL7169 : AC125V)
WARNING
Confirm that the instrument is powered off, and then connect the Power cord.
Connect the Power cord to the instrument first. Cord to be firmly inserted.
Never attempt to make measurement if any abnormal conditions such as abnormal conditions are
noted, such as a broken Cover and exposed metal parts.
When the instrument is not in use, disconnect the Power cord from the outlet.
When unplugging the cord from the mains socket outlet, do so by removing the plug first and not by
pulling the cord.
Power cord connection
Follow the procedure below, and connect the Power cord.
1 Confirm that the instrument is powered off.
2 Connect the Power cord to the Power connector on the instrument.
3 Connect the Power cord plug to the mains socket outlet.
Power supply rating
Rating of power supply is as follows.
Rated supply voltage : 100 to 240V AC (±10%)
Rated power supply frequency : 45 to 65Hz
Max power consumption : 10VA max