4.5 KEW6305
Setting 03” Clamp sensor
Selectable Current range (“Setting 04”) differs by the selected Clamp sensors.
Clamp sensor Current range (” Setting 04 ”)
50A (M-8128) 1 / 5 / 10 / 25 / 50A / AUTO
100A (M-8127) 2 / 10 / 20 / 50 / 100A / AUTO
200A (M-8126) 4 / 20 / 40 / 100 / 200A / AUTO
500A (M-8125) 10 / 50 / 100 / 250 / 500A / AUTO
1000A (M-8124) 20 / 100 / 200 / 500 / 1000A / AUTO
3000A 300 / 1000 / 3000A
Default value (or after system reset) 500A
1 Use the Cursor key on the selection screen, and select “Setting 03”.
2 Press the ENTER key to get the instrument into setting change mode.
3 Present setting (or default value: 500A) blinks. Select the appropriate clamp sensor with the Cursor
key, and then press the ENTER key after making necessary change.
NOTE:
* Accurate results may not be obtained if the Clamp sensors in use do not match the setting
done for sensor.