4.29 KEW6305
Setting 27” V / A range Auto-setting
Following explains how to activate auto-setting for Voltage range (Setting 02), Clamp meter (Setting 03),
Current range (Setting 04).
1 Select the appropriate wiring configuration at Setting 01.
2 Connect the instrument to the circuit under test.
3 Use the Cursor key on the selection screen, and select “Setting 27”.
4 Press the ENTER key to get the instrument into setting change mode.
5 Press the Cursor key to select “ON”, and then press the ENTER key.
When a message “Err” appears on the LCD, please check the connections of Clamp sensors.
NOTE:
* If the instrument fails to detect the connected sensor properly, the default setting (8125 / 500A
Type) will become effective.
* For Current range, “AUTO” will be automatically selected.