4.11 KEW6305
150V 300V 600V 1P2W 1P3W 3P3W 3P4W
5A 10A 20A 50A 100A 200A 500A 1000A
“Setting 09” Recording interval
Following explains how to set the recording interval for integration/ demand measurement.
The recording interval is a time distance to record each measurement data into SD Card or internal
memory.
Setting time
1 / 2 / 5 / 10 / 15 / 20 / 30 sec.,
1 / 2 / 5 / 10 / 15 / 20 / 30 min.,
1 hour
Default value (or after system reset) 30 min.
1 Use the Cursor key on the selection screen, and select “Setting 09”.
2 Press the ENTER key to get the instrument into setting change mode.
3 Previous setting (or default value: 30 min.) blinks. Press the Cursor key to select any desired time,
and then press the ENTER key after making necessary change.
NOTE:
* Selectable interval is limited by the setting done at Setting 16 (Demand measurement cycle).
- An interval greater than the value set at Setting 16 cannot be selected.
- The interval should be divisible by the value set at Setting 16.
- Any of above interval is selectable if “NO” is selected at Setting 16.